BD VACUTAINER? PLASTIC URINE COLLECTOR CUP 364941

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-11-08 for BD VACUTAINER? PLASTIC URINE COLLECTOR CUP 364941 manufactured by Becton, Dickinson & Co..

Event Text Entries

[126660313] Investigation summary: bd had not received samples, but photos were provided by the customer facility for investigation. The photos were evaluated and the customer's indicated failure mode for leakage with the incident lot was not observed. A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product. Investigation conclusion: based on evaluation of the customer photos, the customer? S indicated failure mode for leakage with the incident lot was not observed. The photos did not identify a specific product issue. Root cause description: based on the investigation, a root cause could not be determined. Rationale: complaints received for this device and reported condition will continue to be tracked and trended. Information will be captured on trend reports and monitored. The bd business team regularly reviews the collected data for identification of emerging trends.
Patient Sequence No: 1, Text Type: N, H10


[126660314] It was reported that the bd vacutainer? Plastic urine collector cups were "closing badly" and leaking. There was no report of exposure, injury, or medical intervention.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1917413-2018-03821
MDR Report Key8051824
Date Received2018-11-08
Date of Report2018-11-08
Date of Event2018-06-29
Date Mfgr Received2018-06-29
Device Manufacturer Date2017-09-29
Date Added to Maude2018-11-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactBDX BRETT WILKO
Manufacturer Street9450 SOUTH STATE STREET
Manufacturer CitySANDY UT 84070
Manufacturer CountryUS
Manufacturer Postal84070
Manufacturer Phone8015652845
Manufacturer G1BECTON, DICKINSON & CO.
Manufacturer Street150 SOUTH 1ST AVENUE
Manufacturer CityBROKEN BOW NE 68822
Manufacturer CountryUS
Manufacturer Postal Code68822
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameBD VACUTAINER? PLASTIC URINE COLLECTOR CUP
Generic NameSPECIMEN TRANSPORT AND STORAGE CONTAINER
Product CodeJTW
Date Received2018-11-08
Catalog Number364941
Lot Number7272729
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBECTON, DICKINSON & CO.
Manufacturer Address150 SOUTH 1ST AVENUE BROKEN BOW NE 68822 US 68822


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-11-08

© 2024 FDA.report