VITROS CHEMISTRY PRODUCTS DGXN SLIDES 8343386

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-11-08 for VITROS CHEMISTRY PRODUCTS DGXN SLIDES 8343386 manufactured by Ortho-clinical Diagnostics.

Event Text Entries

[129058186] The investigation determined that lower than expected vitros digoxin (dgxn) results were obtained from a non-vitros mas quality control (qc) fluid and a vitros therapeutic drug monitoring performance verifier (tdm pv) fluid processed using vitros dgxn slides in combination with a vitros 5600 integrated system. The investigation was unable to determine an assignable cause. The initial unexpected results of 1. 77 on (b)(6) 2018 and 1. 98 on (b)(6) 2018 were both resolved by retesting of the control fluid. After a later calibration event, the issue recurred with multiple calibration events unable to resolve it. This indicates the possibilities of improper fluid handling or the vitros dgxn slide lot contributing to the event. However, a vitros dgxn within-run precision test was within the acceptable guidelines indicating vitros dgxn slide lot 1916-0251-2165 was performing as intended. The issue was later resolved by calibrating with a different set of the same lot of calibrators. The vitros 5600 integrated system was performing as intended and did not likely contribute to the event as within-run precision performance was within acceptable guidelines. In addition, continual tracking and trending of complaints has not identified any signals that show a potential systemic issue with vitros dgxn slide lot 1916-0251-2165.
Patient Sequence No: 1, Text Type: N, H10


[129058187] A customer reported lower than expected quality control (qc) results from both a non-vitros mas qc fluid and a vitros therapeutic drug monitoring performance verifier (tdm pv) obtained using vitros digoxin (dgxn) slides on a vitros 5600 integrated system. Mas lot chu1907a l3 results of 1. 77, 1. 98 and 1. 80 ng/ml vs. The expected result of 2. 60 ng/ml. Tdm pv iii r5360 result of 1. 62 ng/ml vs. The expected result of 2. 12 ng/ml. Biased results of the magnitude and direction observed may lead to inappropriate physician action if they were to occur undetected on patient samples. The unexpected vitros dgxn results were generated from non-patient fluids and there was no allegation of patient harm as a result of the event. This report is number 4 of 4 mdr? S for this event. Four (4) 3500a forms are being submitted for this event as 4 devices were involved. This report corresponds to ortho clinical diagnostics inc. Complaint number (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1319809-2018-00123
MDR Report Key8052023
Date Received2018-11-08
Date of Report2018-11-08
Date of Event2018-10-08
Date Mfgr Received2018-10-14
Device Manufacturer Date2018-08-12
Date Added to Maude2018-11-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JAMES A STEVENS
Manufacturer Street100 INDIGO CREEK DRIVE
Manufacturer CityROCHESTER NY 14626
Manufacturer CountryUS
Manufacturer Postal14626
Manufacturer Phone5854533000
Manufacturer G1ORTHO-CLINICAL DIAGNOSTICS
Manufacturer Street513 TECHNOLOGY BLVD.
Manufacturer CityROCHESTER NY 14652
Manufacturer CountryUS
Manufacturer Postal Code14652
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameVITROS CHEMISTRY PRODUCTS DGXN SLIDES
Generic NameIN VITRO DIAGNOSTICS
Product CodeLFM
Date Received2018-11-08
Catalog Number8343386
Lot Number1916-0251-2165
Device Expiration Date2019-01-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerORTHO-CLINICAL DIAGNOSTICS
Manufacturer Address100 INDIGO CREEK DRIVE ROCHESTER NY 14626 US 14626


Patients

Patient NumberTreatmentOutcomeDate
10 2018-11-08

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.