ELECSYS VITAMIN D ASSAY 05894913190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-11-08 for ELECSYS VITAMIN D ASSAY 05894913190 manufactured by Roche Diagnostics.

Event Text Entries

[128420639] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[128420640] The customer received questionable high vitamin d total results for one patient sample from cobas e411 rack analyzer serial number (b)(4). The result was repeatedly >70 ng/ml with a data flag while the calcium result was normal at 7. 8 mg/dl. These results were reported outside of the laboratory. The physician questioned the result as the vitamin d value should be increased or decreased in the same way as the calcium value. The physician suspected an interference with the assay. The customer sent the same sample to another laboratory using the vitamin d total reagent and the result was 90 ng/ml. There was no allegation of an adverse event. Qc results were in the acceptable range.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-04166
MDR Report Key8052426
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-11-08
Date of Report2018-12-10
Date of Event2018-10-23
Date Mfgr Received2018-10-24
Date Added to Maude2018-11-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameELECSYS VITAMIN D ASSAY
Generic NameVITAMIN D TEST SYSTEM
Product CodeMRG
Date Received2018-11-08
Model NumberNA
Catalog Number05894913190
Lot Number34167100
Device Expiration Date2019-06-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-11-08

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.