FUSE FSC-3300-SL-R 60093

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2018-11-08 for FUSE FSC-3300-SL-R 60093 manufactured by Boston Scientific Corporation.

Event Text Entries

[126640529] The exact date of the event is unknown. The provided event date was chosen as a best estimate based on the date that the manufacturer became aware of the event. (b)(4). Although expected, the device has not been received for analysis. Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
Patient Sequence No: 1, Text Type: N, H10


[126640530] It was reported to boston scientific corporation that a fuse 1c colonoscope was used during a procedure performed on an unknown date. According to the complainant, during the procedure, it was noted that the exterior at the tip of the scope was torn. This caused fluids to leak into the scope, resulting in damage to the front camera. There were no patient complications reported as a results of this event. The procedure was completed with another scope. Attempts to obtain additional information regarding the circumstances surrounding this event have been unsuccessful to date. Should additional relevant details become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2018-61220
MDR Report Key8053093
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2018-11-08
Date of Report2019-02-25
Date of Event2018-10-01
Date Mfgr Received2019-01-31
Device Manufacturer Date2015-11-09
Date Added to Maude2018-11-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCAROLE MORLEY
Manufacturer Street300 BOSTON SCIENTIFIC WAY ,
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834015
Manufacturer G1ENDOCHOICE, INC.
Manufacturer Street11810 WILLS ROAD ,
Manufacturer CityALPHARETTA GA 30009
Manufacturer CountryUS
Manufacturer Postal Code30009
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameFUSE
Generic NameENDOSCOPE, ACCESSORIES, IMAGE POST-PROCESSING FOR COLOR ENHANCEMENT
Product CodePEA
Date Received2018-11-08
Returned To Mfg2018-12-07
Model NumberFSC-3300-SL-R
Catalog Number60093
Lot Number0001000378
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer Address300 BOSTON SCIENTIFIC WAY , MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
10 2018-11-08

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.