HELIOS HL3C

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2018-11-09 for HELIOS HL3C manufactured by Kavo Dental Technologies, Llc.

Event Text Entries

[126648486] Upon evaluation by the pelton & crane authorized distributor it was determined the light arm set screw was not properly installed by the distributor during installation. The set screws will prevent the light from unscrewing from the light arm after installation. The pelton & crane installation instructions clearly states to properly install and secure the set screws during installation of the dental light. The installation instructions also lists warnings to ensure the set screws are installed and properly secured. Pelton & crance shipped set screws to the authorized pelton & crane dealer for installation on the dental light.
Patient Sequence No: 1, Text Type: N, H10


[126648487] A dental assistant was cleaning a pelton & crane helios dental light when the light came apart and fell down towards the floor hitting the dental assistant on the side of the head causing a bump. There were no serious injuries reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1017522-2018-00028
MDR Report Key8054586
Report SourceDISTRIBUTOR
Date Received2018-11-09
Date of Report2018-11-09
Date of Event2018-10-15
Date Mfgr Received2018-10-16
Device Manufacturer Date2013-11-26
Date Added to Maude2018-11-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. FRANK RAY
Manufacturer Street11727 FRUEHAUF DRIVE
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal28273
Manufacturer Phone7045877227
Manufacturer G1KAVO DENTAL TECHNOLOGIES, LLC
Manufacturer Street11727 FRUEHAUF DRIVE
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal Code28273
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHELIOS
Generic NameDENTAL LIGHT
Product CodeEAZ
Date Received2018-11-09
Model NumberHL3C
Catalog NumberHL3C
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerKAVO DENTAL TECHNOLOGIES, LLC
Manufacturer Address11727 FRUEHAUF DRIVE CHARLOTTE NC 28273 US 28273


Patients

Patient NumberTreatmentOutcomeDate
10 2018-11-09

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.