MAUDE MDR 8055672

MDR report key
8055672
Report number
2243471-2018-00014
Event key
0
Event type
3
Date of event
2018-09-26
Date received
2018-11-09
Adverse event
3
Product problem
3
Patients in event
0
Reporter occupation
100
Health professional
3
Initial report to FDA
3
Event location
3

Manufacturer Contact#

Contact
NA STACIE-ANN CREIGHTON
Address
1080 US HWY 202 S NA BRANCHBURG NJ 08876 US
Phone
908-908-9082
Report source
M
Manufacturer link flag
Y

Devices#

Seq, Brand, Generic table
SeqBrandGenericManufacturerProduct codeModelCatalogLotPMA510(k)ImplantEvaluatedAvailability
1COBAS 6800 SYSTEMAUTOMATED BLOODBORNE PATHOGEN TEST EQUIPMENTROCHE MOLECULAR SYSTEMS, INC.MZANA05524245001NAR *

Patients#

Sequence, Received, Treatment table
SequenceReceivedTreatmentOutcome
12018-11-090

Event Narratives#

N

Patient 1

PREVIOUS INVESTIGATIVE TESTING SHOWED THAT THE TIGHTNESS CHECK FOR A PROCESSING HEAD WILL FAIL WHEN THE STOP DISK IS VERY LOOSE AND SIGNIFICANTLY LESS THAN 2 NCM TORQUE (RECOMMENDED TORQUE DOCUMENTED IN THE INSTRUMENT SERVICE DOCUMENTATION IS 15 NCM). NO LEAKAGE FROM THE PIPETTE TIPS WAS OBSERVED WHEN THE STOP DISKS WERE TIGHTENED BETWEEN 2 TO 15 NCM USING A TORQUE WRENCH. THE TORQUE MEASUREMENT (FORCE) FOR THE LOOSE STOP DISK ASSOCIATED WITH THE FAILED TIGHTNESS CHECK ON THE CUSTOMER'S COBAS 6800 SYSTEM IS UNKNOWN. THE ERROR FLAGS GENERATED IN THIS CASE COULD BE A RESULT OF THE LEAKAGE CAUSED BY THE LOOSE STOP DISCS. THE ROOT CAUSE INVESTIGATION IS CURRENTLY ON-GOING, AND CORRECTIVE AND PREVENTIVE ACTIONS WILL BE IMPLEMENTED AS APPROPRIATE. (B)(4).

D

Patient 1

A CUSTOMER FROM (B)(6) REPORTED THE GENERATION OF AN INVALID COBAS MPX POSITIVE CONTROL [MPX O (+) CONTROL] DUE TO AN ERROR FLAG (C02H1). WHEN REVIEWING SEVERAL RUNS WITHIN THE CUSTOMER-PROVIDED DATA, IT WAS IDENTIFIED THAT SEVERAL WELLS, CONTAINING SAMPLES OR CONTROLS, HAD THE ERROR CODE P07P, WHICH IS INDICATIVE OF A VOLUME ERROR DURING SUPERNATANT REMOVAL. THE ROCHE FIELD SERVICE ENGINEER VISITED THE CUSTOMER SITE AND FOUND A FAILED RESULT FOR PROCESS HEAD TIGHTNESS CHECK WITH SOME LOOSE STOP DISCS ON THE COBAS 6800 SYSTEM'S PROCESSING TRANSFER HEADS, THAT WERE LIKELY THE CAUSE FOR THE GENERATION OF THE ERROR CODES. THE ROCHE FIELD SERVICE ENGINEER REPLACED ALL STOP DISCS & O-RINGS ON THIS INSTRUMENT PROCESSING TRANSFER HEAD. THERE IS NO ALLEGATION ON DISCREPANT DONOR SAMPLE RESULTS. WHEN ERROR CODES ARE GENERATED FOR THE COBAS MPX CONTROLS, THE TEST RUN IS INVALIDATED AND ALL DONOR SAMPLES INCLUDED IN THE TEST RUN NEED TO BE REPEAT TESTED PER THE INSTRUCTIONS FOR USE. WHEN THE P0P7 ERROR CODE IS GENERATED FOR A DONOR SAMPLE, THE DONOR SAMPLE NEEDS TO BE REPEAT TESTED PER THE INSTRUCTIONS FOR USE. NO HARM OR INJURY WAS REPORTED THROUGH THE CASE.