YASARGIL CLIP 586008

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,foreign report with the FDA on 2018-11-10 for YASARGIL CLIP 586008 manufactured by Medicon E.g..

Event Text Entries

[128627007] The clip brace was slightly bent and had broken near one end. Closing pressure and closing function were found to be within specifications. The bent brace and the traces of blood found on the clip indicate improper handling during the surgical procedure. Apparently force was applied to the clip by another object, first bending and finally breaking the brace.
Patient Sequence No: 1, Text Type: N, H10


[128627008] During a surgical procedure, undue force seems to have been applied to the clip with a foreign object, leading to breakage. The patient did not suffer any harm. No further information has been provided by the distributor, who first indicated that the malfunction occurred during use of the device, and then that it broke during cleaning and sterilization prior to first use. The second statement does not hold, since the returned device showed traces of blood.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8010099-2018-00004
MDR Report Key8058035
Report SourceDISTRIBUTOR,FOREIGN
Date Received2018-11-10
Date of Report2018-11-10
Date Mfgr Received2017-11-15
Device Manufacturer Date2017-10-17
Date Added to Maude2018-11-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactANTON MITTERMUELLER
Manufacturer StreetGAENSAECKER 15
Manufacturer CityTUTTLINGEN, 78532
Manufacturer CountryGM
Manufacturer Postal78532
Manufacturer G1MEDICON E.G.
Manufacturer StreetGAENSAECKER 15
Manufacturer CityTUTTLINGEN, 78532
Manufacturer CountryGM
Manufacturer Postal Code78532
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameYASARGIL CLIP
Generic NameANEURYSM CLIP
Product CodeHCH
Date Received2018-11-10
Catalog Number586008
Lot NumberO314650
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDICON E.G.
Manufacturer AddressGAENSAECKER 15 TUTTLINGEN, 78532 GM 78532


Patients

Patient NumberTreatmentOutcomeDate
10 2018-11-10

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.