ADULT TRACHE DIRECT INTERFACE OPT970

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2018-11-13 for ADULT TRACHE DIRECT INTERFACE OPT970 manufactured by Fisher & Paykel Healthcare Ltd.

Event Text Entries

[128198099] (b)(4). The opt970 interface is used to deliver humidified oxygen to patients via tracheostomy. The interface is held in place by a neck strap and also includes a lanyard which is placed around the patient's neck or attached to the patient's clothing or bedding to remove the load of the breathing circuit from the patient's trache. We are currently in the process of following up with the hospital to retrieve further information and the complaint opt970 device. We will provide a follow up report upon completion of our investigation.
Patient Sequence No: 1, Text Type: N, H10


[128198100] A hospital in (b)(6) reported via a fisher & paykel healthcare (f&p) representative that the opt970 tracheostomy interface disconnected during use. The circuit had come off from the connector part of the opt970. We were further informed that patient was coughing intensely from time to time and it is possible that the opt970 broke as a result of the cough. The hospital informed our field representative that the patient was without humidification at the time of incident but no patient consequences were reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9611451-2018-00971
MDR Report Key8067833
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2018-11-13
Date of Report2018-10-15
Date of Event2018-09-10
Date Mfgr Received2018-10-15
Date Added to Maude2018-11-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. FARANAK GOMAROONI
Manufacturer Street173 TECHNOLOGY DRIVE SUITE 100
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal92618
Manufacturer Phone9194534000
Manufacturer G1FISHER & PAYKEL HEALTHCARE LTD
Manufacturer Street15 MAURICE PAYKEL PLACE EAST TAMAKI
Manufacturer CityAUCKLAND, 2013
Manufacturer CountryNZ
Manufacturer Postal Code2013
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameADULT TRACHE DIRECT INTERFACE
Generic NameBZA
Product CodeBZA
Date Received2018-11-13
Model NumberOPT970
Catalog NumberOPT970
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerFISHER & PAYKEL HEALTHCARE LTD
Manufacturer Address15 MAURICE PAYKEL PLACE EAST TAMAKI AUCKLAND, 2013 NZ 2013


Patients

Patient NumberTreatmentOutcomeDate
10 2018-11-13

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.