TRUE DILATATION CATHETER 0224512

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2018-11-14 for TRUE DILATATION CATHETER 0224512 manufactured by Bard Peripheral Vascular, Inc..

Event Text Entries

[127367173] As the lot number for the device was provided, a manufacturing review will be performed. The sample was not returned to the manufacturer for inspection/evaluation. Therefore, the investigation of the reported event is inconclusive. Based upon the available information, the definitive root cause for this event is unknown. The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device. Upon receipt of new or additional information, a follow-up report will be submitted as applicable. (expiry date: (b)(6) 2020). The information provided by bard represents all of the known information at this time. Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
Patient Sequence No: 1, Text Type: N, H10


[127367174] It was reported that prior to a valvuloplasty procedure, the ptv balloon allegedly ruptured during a test inflation. Another device was used to perform the procedure. There was no reported patient involvement.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2020394-2018-02023
MDR Report Key8071943
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2018-11-14
Date of Report2018-11-14
Date of Event2018-10-16
Date Mfgr Received2018-10-18
Device Manufacturer Date2017-10-23
Date Added to Maude2018-11-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJUDITH LUDWIG
Manufacturer Street1625 W 3RD ST.
Manufacturer CityTEMPE AZ 85281
Manufacturer CountryUS
Manufacturer Postal85281
Manufacturer Phone4803032689
Manufacturer G1C.R. BARD, INC. (GFO)
Manufacturer Street289 BAY ROAD
Manufacturer CityQUEENSBURY NY 12804
Manufacturer CountryUS
Manufacturer Postal Code12804
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRUE DILATATION CATHETER
Generic NameBALLOON VALVULOPLASTY CATHETER
Product CodeOZT
Date Received2018-11-14
Catalog Number0224512
Lot NumberGFBW2983
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBARD PERIPHERAL VASCULAR, INC.
Manufacturer Address1625 W 3RD ST. TEMPE AZ 85281 US 85281


Patients

Patient NumberTreatmentOutcomeDate
10 2018-11-14

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.