NATUS PHOTIC STIM KIT

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2018-11-15 for NATUS PHOTIC STIM KIT manufactured by Natus Medical Inc. Dba Excle-tech Ltd (xltek).

Event Text Entries

[128403747] Following natus complaint procedure, the photic system was investigated. The cable found broken at the end of lower arm. Normal wear and tear failure due to mechanical movement.
Patient Sequence No: 1, Text Type: N, H10


[128403748] During photic activation the power cable starting sparking. Photic power cable had exposed wires after sparks occured. No one was hurt.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9612330-2018-00018
MDR Report Key8074454
Report SourceHEALTH PROFESSIONAL
Date Received2018-11-15
Date of Report2016-02-13
Date of Event2016-02-13
Date Mfgr Received2016-02-13
Date Added to Maude2018-11-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. SANJAY MEHTA
Manufacturer Street2568 BRISTOL CIRCLE
Manufacturer CityOAKVILLE, ONTARIO L6H5S1
Manufacturer CountryCA
Manufacturer PostalL6H5S1
Manufacturer G1NATUS MEDICAL INCORPORATED DBA EXCEL-TECH (XLTEK)
Manufacturer Street2568 BRISTOL CIRCLE
Manufacturer CityOAKVILLE, ONTARIO L6H5S1
Manufacturer CountryCA
Manufacturer Postal CodeL6H5S1
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameNATUS PHOTIC STIM KIT
Generic NamePHOTIC STIMULATOR
Product CodeGWE
Date Received2018-11-15
Returned To Mfg2016-03-03
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerNATUS MEDICAL INC. DBA EXCLE-TECH LTD (XLTEK)
Manufacturer Address2568 BRISTOL CIRCLE OAKVILLE, OAKVILLE ONTARIO L6H 5S1 CA L6H 5S1


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-11-15