MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-11-15 for BARD CARDIOTHORACIC ADAPTOR 072510 manufactured by C.r. Bard, Inc. (covington) -1018233.
[127549794]
The investigation is still in progress. Once the investigation is complete a supplemental report will be filed. The device was not returned.
Patient Sequence No: 1, Text Type: N, H10
[127549795]
It was reported that two of the ports of the 3-in-1 drain connector became unplugged during use, which introduced air into the thoracic cavity. There was no reported injury to the patient, however the complainant noted that there was a risk of respiratory compromise. The connector was being used with a single drain, so the ports that became disconnected were not in use.
Patient Sequence No: 1, Text Type: D, B5
| Report Number | 1018233-2018-05449 |
| MDR Report Key | 8075796 |
| Date Received | 2018-11-15 |
| Date of Report | 2018-12-12 |
| Date Mfgr Received | 2018-11-26 |
| Date Added to Maude | 2018-11-15 |
| Event Key | 0 |
| Report Source Code | Manufacturer report |
| Manufacturer Link | Y |
| Number of Patients in Event | 0 |
| Adverse Event Flag | 3 |
| Product Problem Flag | 3 |
| Reprocessed and Reused Flag | 3 |
| Health Professional | 3 |
| Initial Report to FDA | 3 |
| Report to FDA | 3 |
| Event Location | 3 |
| Manufacturer Contact | YONIC ANDERSON |
| Manufacturer Street | 8195 INDUSTRIAL BLVD |
| Manufacturer City | COVINGTON GA 30014 |
| Manufacturer Country | US |
| Manufacturer Postal | 30014 |
| Manufacturer Phone | 7707846100 |
| Manufacturer G1 | C.R. BARD, INC. (COVINGTON) -1018233 |
| Manufacturer Street | 8195 INDUSTRIAL BLVD |
| Manufacturer City | COVINGTON GA 30014 |
| Manufacturer Country | US |
| Manufacturer Postal Code | 30014 |
| Single Use | 3 |
| Previous Use Code | 3 |
| Event Type | 3 |
| Type of Report | 0 |
| Brand Name | BARD CARDIOTHORACIC ADAPTOR |
| Generic Name | DRAIN CONNECTOR |
| Product Code | GCE |
| Date Received | 2018-11-15 |
| Catalog Number | 072510 |
| Lot Number | UNK |
| Operator | HEALTH PROFESSIONAL |
| Device Availability | N |
| Device Age | DA |
| Device Eval'ed by Mfgr | R |
| Device Sequence No | 1 |
| Device Event Key | 0 |
| Manufacturer | C.R. BARD, INC. (COVINGTON) -1018233 |
| Manufacturer Address | 8195 INDUSTRIAL BLVD COVINGTON GA 30014 US 30014 |
| Patient Number | Treatment | Outcome | Date |
|---|---|---|---|
| 1 | 0 | 2018-11-15 |