30-70ML CELL FREEZE CRYOGENIC STORAGE CONTAINER/3 LEADS CF-250-3L

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2018-11-16 for 30-70ML CELL FREEZE CRYOGENIC STORAGE CONTAINER/3 LEADS CF-250-3L manufactured by Charter Medical, Ltd..

Event Text Entries

[127866609] The sample was not returned for investigation, however the user provided pictures of a bag from lot 149037. The pictures indicated a bag with irregular film cracks over the top portion of the bag and into the port area. No other issues were observed from the bag pictures. A capa had previously been issued to investigate film breaks of this nature. A review of the capa investigation indicated failure modes of this type have been reproduced by introducing an air bubble inside the bag prior to freezing. An air bubble could rapidly expand during the thawing process, resulting in a film tear. The evidence observed from the provided pictures suggests the likely root cause of the cracked film was the existence of an air bubble in the bag prior to freezing. Precautions for air bubbles inside the bag prior to freezing are addressed in the product ifu's.
Patient Sequence No: 1, Text Type: N, H10


[127866610] The information provided by the customer indicated a bag cracked during the thawing process. The crack occurred in a laboratory setting and did not involve a patient. There was no report of an adverse event or need for additional medical intervention as a result of this occurrence.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1066733-2018-00021
MDR Report Key8077389
Report SourceUSER FACILITY
Date Received2018-11-16
Date of Report2018-11-16
Date of Event2018-10-17
Date Mfgr Received2018-10-17
Device Manufacturer Date2017-12-07
Date Added to Maude2018-11-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. TODD MEINECKE
Manufacturer Street3948-A WESTPOINT BLVD.
Manufacturer CityWINSTON SALEM NC 27103
Manufacturer CountryUS
Manufacturer Postal27103
Manufacturer Phone3367686447
Manufacturer G1CHARTER MEDICAL, LTD.
Manufacturer Street3948-A WESTPOINT BLVD.
Manufacturer CityWINSTON SALEM NC 27103
Manufacturer CountryUS
Manufacturer Postal Code27103
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name30-70ML CELL FREEZE CRYOGENIC STORAGE CONTAINER/3 LEADS
Generic Name30-70ML CELL FREEZE CRYOGENIC STORAGE CONTAINER/3 LEADS
Product CodeLPZ
Date Received2018-11-16
Model NumberCF-250-3L
Catalog NumberCF-250-3L
Lot Number149037
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCHARTER MEDICAL, LTD.
Manufacturer Address3948-A WESTPOINT BLVD. WINSTON SALEM NC 27103 US 27103


Patients

Patient NumberTreatmentOutcomeDate
10 2018-11-16