CIDEX? OPA TEST STRIPS 20392

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2018-11-16 for CIDEX? OPA TEST STRIPS 20392 manufactured by Advanced Sterilization Products.

Event Text Entries

[129158586] Asp complaint ref #: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[129158587] A customer reported using expired cidex? Opa test strips to test their cidex? Opa solution and the instruments that were cleaned in the tested cidex? Opa solution were released and used on patients. They stated they used the test strips because the instructions for use (ifu) states the strips can be used for 90-days after opening the bottle, and the cidex? Opa strips came out positive (or purple) indicating the cidex? Opa solution meets the minimum effective concentration (mec). There was no report of infection, injury or harm to patient(s) associated with this issue. The customer was advised to discard any remaining cidex? Opa test strips and to follow the instructions for use (ifu) which states the unused cidex? Opa test strips should be discarded after the expiration date on the outside of the bottle even if it has not reached the 90-day open date. As a matter of policy, advanced sterilization products (asp) has decided to report cases where a customer uses expired product and releases it for use on patients since they cannot guarantee high-level disinfection has been achieved.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2084725-2018-00763
MDR Report Key8078374
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2018-11-16
Date of Report2018-11-02
Date of Event2018-11-02
Date Mfgr Received2019-01-21
Device Manufacturer Date2017-10-01
Date Added to Maude2018-11-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. GABRIEL ALFAGEME
Manufacturer Street33 TECHNOLOGY DRIVE
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal92618
Manufacturer Phone949789-868
Manufacturer G1ADVANCED STERILIZATION PRODUCTS
Manufacturer Street33 TECHNOLOGY DRIVE
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal Code92618
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCIDEX? OPA TEST STRIPS
Generic NameBIOCIDES TEST STRIPS
Product CodeJOJ
Date Received2018-11-16
Catalog Number20392
Lot Number031779
Device Expiration Date2018-08-28
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerADVANCED STERILIZATION PRODUCTS
Manufacturer Address33 TECHNOLOGY DRIVE IRVINE CA 92618 US 92618


Patients

Patient NumberTreatmentOutcomeDate
10 2018-11-16

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.