PROTECTION PLUS AIRSTREAM/BREATHABLE UNDERPAD MSC282110

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2007-01-12 for PROTECTION PLUS AIRSTREAM/BREATHABLE UNDERPAD MSC282110 manufactured by Medline Industries, Inc..

Event Text Entries

[582278] Nurse opened the medline protection plus airstream/breathable underpad, which is vacuum packed. When the pack opened, particles of the absorbent flew into the air which got into the nurse's eyes as well as was inhaled. Further analysis of this event revealed the following: after opening the package, it is necessary to unfold each pad before use. If you take the time to pull the folds out slowly and gently, the pad seems fine. If instead, you snap it out quickly to open the folds, the seal around the underpad will break. After the seal is compromised, the pad will separate and all of the absorbent material will become airborne. The absorbent will break apart into several clumps, as well as thousands of very fine particles, which will float into the air. This is not only an eye irritant, but also represents a much worse respiratory hazard for both the healthcare worker and patients. These particular pads are currently being removed from our facility.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number808277
MDR Report Key808277
Date Received2007-01-12
Date of Report2007-01-12
Date of Event2007-01-10
Report Date2007-01-12
Date Reported to FDA2007-01-12
Date Added to Maude2007-01-25
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NamePROTECTION PLUS AIRSTREAM/BREATHABLE UNDERPAD
Generic NameUNDERPAD, PROTECTION
Product CodeLWG
Date Received2007-01-12
Model NumberMSC282110
Catalog NumberMSC282110
Lot NumberNA
ID Number*
OperatorNURSE
Device AvailabilityN
Device Age1 DY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key795762
ManufacturerMEDLINE INDUSTRIES, INC.
Manufacturer Address1170 NORTHPOINT BLVD. WAUKEGAN IL 60085 US


Patients

Patient NumberTreatmentOutcomeDate
10 2007-01-12

© 2026 FDA.report