LEOPARD IMPLANT 5 DEGREE LORDOTIC 12MM 186448012

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2006-10-19 for LEOPARD IMPLANT 5 DEGREE LORDOTIC 12MM 186448012 manufactured by Depuy Spine, Inc..

Event Text Entries

[605292] Contact reported that the leopard implant was inserted with the inserter. While being impacted into position with the straight tamp, a small position of the implant broke off. The fragment was removed and the surgeon left the implant in place. There was no adverse patient consequence, however, as a compromised implant was left in place an mdr is being filed to document this event.
Patient Sequence No: 1, Text Type: D, B5


[7820435] No conclusions can be made at this time. The most likely cause of the event is excessive force placed on the cage during insertion. This type of event is not unanticipated and the instructions for use (ifu) supplied with the device warns against the use of excessive force.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1526439-2006-00227
MDR Report Key808364
Report Source07
Date Received2006-10-19
Date of Event2006-09-19
Date Mfgr Received2006-09-19
Date Added to Maude2007-01-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactMATTHEW KING
Manufacturer Street325 PARAMOUNT DRIVE
Manufacturer CityRAYNHAM MA 02767
Manufacturer CountryUS
Manufacturer Postal02767
Manufacturer Phone5088283106
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameLEOPARD IMPLANT 5 DEGREE LORDOTIC 12MM
Generic NameINTERBODY DEVICE
Product CodeMRX
Date Received2006-10-19
Returned To Mfg2006-10-18
Model NumberNA
Catalog Number186448012
Lot NumberUNK
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Implant FlagY
Date RemovedA
Device Sequence No1
Device Event Key795848
ManufacturerDEPUY SPINE, INC.
Manufacturer Address325 PARAMOUNT DR. RAYNHAM MA 02767 US
Baseline Brand NameLEOPARD IMPLANT 5 DEGREE LORDOTIC, 12MM
Baseline Generic NameVBR IMPLANT
Baseline Model NoNA
Baseline Catalog No186448012
Baseline IDNA


Patients

Patient NumberTreatmentOutcomeDate
10 2006-10-19