[129184000] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[129184001] The customer received a questionable high elecsys total psa immunoassay result for one patient sample from cobas e 801 module 1617-07. The result did not fit with the clinical picture for the patient. The patient was being treated for prostate cancer and the previous total psa results had been low. Refer to the attachment to the medwatch for all patient data. On (b)(6) 2018, the total psa result was 2. 7 ng/ml. A few days later, a new sample was drawn from the patient and the total psa result was 0. 27 ng/ml. There was no allegation of an adverse event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-04378
MDR Report Key8085288
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-11-19
Date of Report2019-01-04
Date of Event2018-10-18
Date Mfgr Received2018-11-01
Date Added to Maude2018-11-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameELECSYS TOTAL PSA IMMUNOASSAY
Generic NameTOTAL,PROSTATE SPECIFIC ANTIGEN(NONCOMPLEXED&COMPLEXED)
Product CodeMTF
Date Received2018-11-19
Model NumberNA
Catalog Number04491734190
Lot NumberASKU
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-11-19