146-S MAPLE LEAF, SMALL

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2018-11-20 for 146-S MAPLE LEAF, SMALL manufactured by Becker Orthopedic Appliance Co.

Event Text Entries

[127877939] A review of all return, compliant, and feedback information on the maple leaf? For the past seven years was conducted and no other such narrative was documented. There were zero instances of pressure ulcers, tightness, or fit issues in any of the other cases of the maple leaf being used. I was "instructued" by (b)(6) to add a narrative that this submission was not within 5 business days of receipt due to our firm being in the enrollment process of electronic submission. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[127877940] A maple leaf? (146-s) was fitted on a patient on, or around, (b)(6) 2018 at (b)(6) hospital (b)(6). The orthotic brace was adjusted for range of motion, rotation, and flexion on, or around, (b)(6) 2018, (b)(6) 2018, and (b)(6) 2018. On (b)(6) 2018, a pressure ulcer was noted on the back edge of the right thigh cuff of the maple leaf?. The patient was wearing the orthotic brace continuously from the fitting of the brace until the discovery of the pressure ulcer. The patient was fitted with the orthotic brace to stabilize a repaired fracture and was in multiple laying, supine, prone, and recumbent positions during their inpatient and outpatient stays.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1824252-2018-00001
MDR Report Key8089044
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2018-11-20
Date Mfgr Received2018-10-23
Date Added to Maude2018-11-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactERICA GUICE
Manufacturer Street635 EXECUTIVE DR
Manufacturer CityTROY MI 48083
Manufacturer CountryUS
Manufacturer Postal48083
Manufacturer Phone2485887480
Manufacturer G1BECKER ORTHOPEDIC APPLIANCE CO
Manufacturer Street635 EXECTUIVE DR
Manufacturer CityTROY MI 48083
Manufacturer CountryUS
Manufacturer Postal Code48083
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name146-S MAPLE LEAF, SMALL
Generic NameHIP ABDUCTION ORTHOSIS
Product CodeITS
Date Received2018-11-20
Model Number146-S
Catalog Number146-S
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBECKER ORTHOPEDIC APPLIANCE CO
Manufacturer Address635 EXECTUIVE DR TROY MI 48083 US 48083


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-11-20

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.