ETEST? VANCOMYCIN VA 256 WW S30 412488

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-11-20 for ETEST? VANCOMYCIN VA 256 WW S30 412488 manufactured by Biomerieux, Sa.

Event Text Entries

[129350070] A customer in (b)(6) reported a false susceptible result for an enterococcus faecium vanb positive strain in association with etest? Vancomycin va 256 ww s30. The customer stated they obtained discrepant results for vancomycin between etest and the vitek? 2 ast-p655 card. The etest initial and repeat results were vancomycin susceptible (mic=-2), and the ast-655 card was initially susceptible (mic=0. 5), but changed to resistant as the vana phenotype was detected. The strain was tested by pcr vanb twice and was positive, confirming a vanb positive phenotype. A vanb enterococcus faecium strain is expected to be vancomycin resistant. The customer did not report any patient information. There is no indication or report from the hospital or treating physician to biom? Rieux that the discrepant result led to any adverse event related to the patient's state of health. A biom? Rieux internal investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2018-00244
MDR Report Key8089293
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-11-20
Date of Report2018-11-20
Date Mfgr Received2018-10-25
Date Added to Maude2018-11-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. DEBRA BROYLES
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer G1BIOMERIEUX, SA
Manufacturer Street3 ROUTE DE PORT MICHAUD
Manufacturer CityLA BALME, LES GROTTES ISERE 38390
Manufacturer CountryFR
Manufacturer Postal Code38390
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameETEST? VANCOMYCIN VA 256 WW S30
Generic NameETEST? VANCOMYCIN VA 256 WW S30
Product CodeJWY
Date Received2018-11-20
Catalog Number412488
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX, SA
Manufacturer Address3 ROUTE DE PORT MICHAUD LA BALME, LES GROTTES ISERE 38390 FR 38390


Patients

Patient NumberTreatmentOutcomeDate
10 2018-11-20

© 2026 FDA.report