MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2018-11-20 for CLINICAL CHEMISTRY GLUCOSE 03L82-21 manufactured by Abbott Manufacturing Inc.
[127944592]
All available patient information was included. Additional patient details are not available. An evaluation is in process. A final report will be submitted when the evaluation is complete.
Patient Sequence No: 1, Text Type: N, H10
[127944593]
The customer observed falsely elevated glucose results on the architect c16000 analyzer. The following data was provided: t0 169. 1 mg/dl 92. 1 mg/dl; t1 240. 3 mg/dl 144 mg/dl; t2 216 mg/dl, 101. 2 mg/dl. Due to imprecise glucose results, the doctor started a treatment based on the first result, which turned out to be inappropriate, causing significant hypoglycemia in the patient. No further details regarding the treatment or how the patient is doing, are yet available.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1628664-2018-02026 |
MDR Report Key | 8090294 |
Report Source | FOREIGN,HEALTH PROFESSIONAL |
Date Received | 2018-11-20 |
Date of Report | 2019-02-04 |
Date of Event | 2018-11-03 |
Date Mfgr Received | 2019-01-09 |
Date Added to Maude | 2018-11-20 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | NOEMI ROMERO-KONDOS, RN BSN |
Manufacturer Street | 100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3 |
Manufacturer City | ABBOTT PARK IL 600643537 |
Manufacturer Country | US |
Manufacturer Postal | 600643537 |
Manufacturer Phone | 224667-512 |
Manufacturer G1 | ABBOTT MANUFACTURING INC |
Manufacturer Street | 1921 HURD DRIVE |
Manufacturer City | IRVING TX 75038 |
Manufacturer Country | US |
Manufacturer Postal Code | 75038 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | CLINICAL CHEMISTRY GLUCOSE |
Generic Name | GLUCOSE |
Product Code | CFR |
Date Received | 2018-11-20 |
Catalog Number | 03L82-21 |
Lot Number | UNKNOWN |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ABBOTT MANUFACTURING INC |
Manufacturer Address | 1921 HURD DRIVE IRVING TX 75038 US 75038 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other | 2018-11-20 |