GLUMA DESENSITIZER POWERGEL 66043451

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign report with the FDA on 2018-11-21 for GLUMA DESENSITIZER POWERGEL 66043451 manufactured by Kulzer, Gmbh.

Event Text Entries

[128097483] (b)(4). Although we have established that user error caused the injuries to the patients and the patient did not seek treatment from a secondary provider, and all tissues have healthed without permanent damage, kulzer, llc is reporting this out of an abundance of caution, due to the (b)(6) requesting that this be reported and at the request of our global safety officer.
Patient Sequence No: 1, Text Type: N, H10


[128097484] Incident occured in (b)(6). (b)(6) yo female. Hyposalivation, smoker. Patient medications include lamictal and somac. After treatment with gluma desensitizer powergel, the patient had gingival injuries, headache, skin reaction, burning feeling, swelling/edema. Treating dds applied the product to what appears to be multiple areas. The dds did not use isolation, nor did they rinse the material after application. The product was applied and the patient was allowed to leave the office with the material still on the teeth. Per the office on (b)(6) 2018, the patient did not seek secondary treatment and they had been seen back at the office for a follow up visit and the provider was able to confirm that the injury has completely healed without permanent damage.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610902-2018-00005
MDR Report Key8096346
Report SourceFOREIGN
Date Received2018-11-21
Date of Report2018-11-20
Date of Event2018-10-24
Date Mfgr Received2018-10-25
Date Added to Maude2018-11-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS AMBER BROWN
Manufacturer Street4315 S. LAFAYETTE BLVD
Manufacturer CitySOUTH BEND IN 46614
Manufacturer CountryUS
Manufacturer Postal46614
Manufacturer Phone5472995411
Manufacturer G1KULZER GMBH
Manufacturer StreetPHILIPP-REIS-STRA?E 8/13
Manufacturer CityWEHRHEIM, D-61273
Manufacturer CountryGM
Manufacturer Postal CodeD-61273
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameGLUMA DESENSITIZER POWERGEL
Generic NameDESENSITIZING AGENT
Product CodeKLE
Date Received2018-11-21
Catalog Number66043451
Lot NumberUNKNOWN
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerKULZER, GMBH
Manufacturer AddressPHILIPP-REIS-STRASSE 8/13 WEHRHEIM, D-61273 GM D-61273


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-11-21