NEWBORN BILI-BONNET 01B

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2007-01-29 for NEWBORN BILI-BONNET 01B manufactured by Small Beginnings Inc..

Event Text Entries

[20161440] In newborn nursery a full-term female born via vaginal delivery with apgars nine and nine noted to be jaundiced was started on phototherapy. Small beginnings newborn bili-bonnet phototherapy mask was placed over the baby's eyes. At ten hours of age the baby was found not breathing and the mask was found positioned below the eyes. Code was called and baby was resuscitated and transferred to neonatal intensive care unit. Baby continued on life support. Evidence of worsening hypoxic-ischemic encephalopathy. After extensive discussion with family, life support removed. Autopsy pending.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number810312
MDR Report Key810312
Date Received2007-01-29
Date of Report2007-01-29
Date of Event2006-12-31
Report Date2007-01-29
Date Reported to FDA2007-01-29
Date Added to Maude2007-01-31
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameNEWBORN BILI-BONNET
Generic NamePHOTOTHERAPY, MASK
Product CodeFOK
Date Received2007-01-29
Model Number01B
Catalog Number01B
Lot Number07312006FB
ID Number*
Device AvailabilityN
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key797795
ManufacturerSMALL BEGINNINGS INC.
Manufacturer Address17525 ALDER STREET, SUITE 28 HESPERIA CA 92345 US


Patients

Patient NumberTreatmentOutcomeDate
101. Death 2007-01-29

© 2026 FDA.report