NEEDLE 1/2 CIRCLE TAPER POINT MAYO CATGUT .050X1.260 STERILE 216704

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-11-26 for NEEDLE 1/2 CIRCLE TAPER POINT MAYO CATGUT .050X1.260 STERILE 216704 manufactured by Aspen Surgical Products, Caledonia.

Event Text Entries

[129322966] Aspen surgical received a report from the end user indicating that a circle taper point needle broke during a procedure. The actual device was not available for evaluation. No photo was provided for review. Manufacturing lot number was available. The end user indicated that during a right open hamstring repair the needle tip broke. The surgeon completed the surgery and xrays were not performed. The broken piece was not recovered or found. The customer indicated the needle tip may not be in the patient, and the surgeon did not pursue any attempts to locate the needle tip if inside the patient. There was no injury alleged or prolonged procedure to retrieve. The dhr and analysis results were reviewed. All samples passed acceptance criteria. No non-conformance's were noted related to the reported issue. According to aspen surgical's ifu, the needles should not be clamped directly on the eye of the needle to the point and should be applied to the flat portion of the needle about 1/4 of the needle length from the eye end to prevent needle breaks. Based on this information, due to no picture or sample being available to aid with the complaint investigation, no further action is required. The cause could not be established for needle breakage.
Patient Sequence No: 1, Text Type: N, H10


[129322967] Aspen surgical received a report from the end user indicating that a circle taper point needle broke during a procedure. This report was filed in our complaint handling system as complaint (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1836161-2018-00122
MDR Report Key8103256
Date Received2018-11-26
Date of Report2018-11-01
Date of Event2018-10-31
Date Mfgr Received2018-11-01
Device Manufacturer Date2018-02-26
Date Added to Maude2018-11-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JORDAN HACKERT
Manufacturer Street6945 SOUTHBELT DR. S.E.
Manufacturer CityCALEDONIA MI 49316
Manufacturer CountryUS
Manufacturer Postal49316
Manufacturer Phone6165367508
Manufacturer G1SAME AS ABOVE
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameNEEDLE 1/2 CIRCLE TAPER POINT MAYO CATGUT .050X1.260 STERILE
Generic NameSURGICAL NEEDLE
Product CodeGAB
Date Received2018-11-26
Model Number216704
Lot Number150260
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerASPEN SURGICAL PRODUCTS, CALEDONIA
Manufacturer Address6945 SOUTHBELT DR. S.E. CALEDONIA MI 49316 US 49316


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-11-26