PREVI? COLOR GRAM 29551R

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2018-11-27 for PREVI? COLOR GRAM 29551R manufactured by Biomerieux, Sa.

Event Text Entries

[129580608] A customer in the united states contacted biom? Rieux to report observing crystals in the gram stains prepared by the previ? Color gram instrument. The customer stated the crystals look like gram negative rods, and have caused some laboratory technicians to misinterpret the result. Manual gram staining does not create the reported effect. The customer stated there have been a few incidences of calling gram negative rods (gnrs) on spinal fluids. These situations cause a delay (duration not disclosed) due to further testing via manual staining. The customer stated there have been no treatment decisions by a physician based on a discrepant gram stain result. Biom? Rieux investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2018-00245
MDR Report Key8105626
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2018-11-27
Date of Report2018-11-27
Date Mfgr Received2018-10-28
Date Added to Maude2018-11-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JEFF SCANLAN
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147318694
Manufacturer G1BIOMERIEUX, SA
Manufacturer Street3 ROUTE DE PORT MICHAUD
Manufacturer CityLA BALME, LES GROTTES ISERE 38390
Manufacturer CountryFR
Manufacturer Postal Code38390
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePREVI? COLOR GRAM
Generic NamePREVI? COLOR GRAM
Product CodeKPA
Date Received2018-11-27
Model Number29551R
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX, SA
Manufacturer Address3 ROUTE DE PORT MICHAUD LA BALME, LES GROTTES ISERE 38390 FR 38390


Patients

Patient NumberTreatmentOutcomeDate
10 2018-11-27

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.