VE HDTV INTERNATIONAL KIT 0240030960I

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2018-11-27 for VE HDTV INTERNATIONAL KIT 0240030960I manufactured by Stryker Endoscopy-san Jose.

Event Text Entries

[128460762] Alleged failure: displayed color stripes. Confirmed failure: damaged cma, main board failure, scratched pp filter. Probable root cause: scaler/descaler board. Power supply. Ac inlet board. Connectors. Firmware. Switch board. Use error. The product was returned for investigation and the failure mode will be monitored for future reoccurrence. The device manufacture date is not known.
Patient Sequence No: 1, Text Type: N, H10


[128460763] It was reported that stripe colors were displayed on the monitor, which resulted to conversion to open surgery.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0002936485-2018-00704
MDR Report Key8107437
Report SourceCOMPANY REPRESENTATIVE
Date Received2018-11-27
Date of Report2018-11-27
Date of Event2018-07-07
Date Mfgr Received2018-11-01
Date Added to Maude2018-11-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. MICHELLE TO
Manufacturer Street5900 OPTICAL COURT
Manufacturer CitySAN JOSE CA 95138
Manufacturer CountryUS
Manufacturer Postal95138
Manufacturer Phone4087542000
Manufacturer G1STRYKER ENDOSCOPY-SAN JOSE
Manufacturer Street5900 OPTICAL COURT
Manufacturer CitySAN JOSE CA 95138
Manufacturer CountryUS
Manufacturer Postal Code95138
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVE HDTV INTERNATIONAL KIT
Generic NameCAMERA, TELEVISION, SURGICAL, WITHOUT AUDIO
Product CodeFWB
Date Received2018-11-27
Returned To Mfg2018-10-26
Catalog Number0240030960I
Lot NumberVEH263D0281
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER ENDOSCOPY-SAN JOSE
Manufacturer Address5900 OPTICAL COURT SAN JOSE CA 95138 US 95138


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-11-27