CUP BIOPSY FORCEPS 220130

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-11-28 for CUP BIOPSY FORCEPS 220130 manufactured by Cook Inc.

Event Text Entries

[128740986] The field for section could not be populated due to a technical limitation. Section common name: endoscopic grasping/cutting instrument, non-powered section product code: ocz. This mdr is being filed after the associated complaint was reviewed under remediation protocol (b)(4), complaint/mdr retrospective review and remediation and reassessed as reportable. Additional complaint investigation and record remediation was not performed.
Patient Sequence No: 1, Text Type: N, H10


[128740987] It was reported that during the biopsy in flexible ureteroscopy procedure, the sheath (outer layer) of the forceps separated completely from the metallic part (coil) of the sheath. This occurred inside the scope and outside of the patient. It was further noted that the small scope had scraped the sheath off the coil assembly. There were no adverse effects to the patient reported due to the event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1820334-2018-03569
MDR Report Key8112898
Date Received2018-11-28
Date of Report2018-11-28
Date of Event2016-02-25
Date Mfgr Received2018-11-26
Date Added to Maude2018-11-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. LARRY POOL
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal47404
Manufacturer Phone8128294891
Manufacturer G1COOK INC
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal Code47404
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCUP BIOPSY FORCEPS
Product CodeOCZ
Date Received2018-11-28
Returned To Mfg2016-03-15
Model NumberNA
Catalog Number220130
Lot NumberUNKNOWN
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCOOK INC
Manufacturer Address750 DANIELS WAY BLOOMINGTON IN 47404 US 47404


Patients

Patient NumberTreatmentOutcomeDate
10 2018-11-28

© 2026 FDA.report