STATSPIN SSVT-1 CENTRIFUGE X00-003917-001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2018-11-28 for STATSPIN SSVT-1 CENTRIFUGE X00-003917-001 manufactured by Beckman Coulter.

Event Text Entries

[128924967] The customer acknowledge they failed to follow instructions to use the plastic inner shield during spin cycles as required. Customer technical support (cts) provided customer with most recent statspin safety notice, safety shield and a trained customer with importance of installing shield. The customer confirmed there was no damage to lid, hinge or latch of centrifuge. Cts initiated return merchandise authorization (rma) process to return centrifuge for repair. The customer was provided with a replacement centrifuge to resolve the issue. Beckman coulter internal identifier is case- (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[128924968] The customer reported statspin ssvt-1 centrifuge rotor broke during use and broken rotor pieces hit a lab associate in the leg. There was no adverse event reported. The associate did not seek medical attention for the incident.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2050012-2018-01039
MDR Report Key8113289
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2018-11-28
Date of Report2018-11-28
Date of Event2018-10-22
Date Facility Aware2018-11-16
Date Mfgr Received2018-11-16
Device Manufacturer Date2012-10-22
Date Added to Maude2018-11-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR DAVID DAVIS
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 928218000
Manufacturer CountryUS
Manufacturer Postal928218000
Manufacturer Phone7149613796
Manufacturer G1BECKMAN COULTER
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 928218000
Manufacturer CountryUS
Manufacturer Postal Code928218000
Single Use3
Previous Use Code3
Removal Correction NumberFA-24395
Event Type3
Type of Report3

Device Details

Brand NameSTATSPIN SSVT-1 CENTRIFUGE
Generic NameCENTRIFUGES (MICRO, ULTRA, REFRIGERATED) FOR CLINICAL USE
Product CodeJQC
Date Received2018-11-28
Model NumberSSVT-1 CENTRIFUGE
Catalog NumberX00-003917-001
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER
Manufacturer Address250 S. KRAEMER BLVD. BREA CA 928218000 US 928218000


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-11-28

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.