ECHOSCREEN TEOAE HEARING SCREENER 010128

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2007-01-10 for ECHOSCREEN TEOAE HEARING SCREENER 010128 manufactured by Fischer-zoth Diagnosesysteme, Gmbh.

Event Text Entries

[604294] During the charge cycle, the battery pack of the echoscreen teoae hearing screener caught on fire at hospital. The estimated charge time before the incident occurred in 2-3 hours. The battery charger was plugged into a wall outlet or power strip. There was a power surge protector from the power source being used. There were no power supply interruptions during the battery charging cycle. There was no liquid spilled in the area where the battery was being charged. There were no objects placed on top of the battery during the charging cycle. The battery was not exposed to mechanical shock, nor was it damaged mechanically. It did not fall to the floor. The indicator light was functioning properly (flashing slowly) during the charge cycle.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2938854-2007-00001
MDR Report Key811481
Report Source05,06
Date Received2007-01-10
Date of Report2007-01-10
Date of Event2006-12-06
Date Mfgr Received2006-12-11
Date Added to Maude2007-02-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactLUANNE NG, MGR
Manufacturer Street1501 INDUSTRIAL RD
Manufacturer CitySAN CARLOS CA 94070
Manufacturer CountryUS
Manufacturer Postal94070
Manufacturer Phone8002553901
Manufacturer G1FISCHER-ZOTH DIAGOSESYSTEME, GMBH
Manufacturer Street*
Manufacturer CityGERMERING
Manufacturer CountryGM
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameECHOSCREEN TEOAE HEARING SCREENER
Generic NameHEARING SCREENER
Product CodeEWO
Date Received2007-01-10
Returned To Mfg2006-12-11
Model NumberTEOAE
Catalog Number010128
Lot NumberBATTERY - 050605
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key798967
ManufacturerFISCHER-ZOTH DIAGNOSESYSTEME, GMBH
Manufacturer AddressWALTER-KOLBENHOFF-STR 34 GERMERING GM
Baseline Brand NameECHOSCREEN TEROAE HEARING SCREENER
Baseline Generic NameHEARING SCREENER
Baseline Model NoTEOAE
Baseline Catalog No010128
Baseline ID*


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2007-01-10