ION TORRENT? ONCOMINE LUNG CELL-FREE TOTAL NUCLEIC ACID (CFTNA) RESEARCH ASSAY A35864

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-12-01 for ION TORRENT? ONCOMINE LUNG CELL-FREE TOTAL NUCLEIC ACID (CFTNA) RESEARCH ASSAY A35864 manufactured by Life Technologies Corporation.

Event Text Entries

[129753557] The ion torrent? Oncomine lung cell-free total nucleic acid (cftna) research assay is labeled "for research use only. Not for use in diagnostic procedures. " (ruo) no patient impact at the biocept site was reported, however, the manufacturer, thermo fisher scientific, is submitting this mdr due to a possible false positive result` for the t790m variant at the (b)(6) if the lab utilizes the ruo product to assess patient samples. The unit in question was not returned by the customer. However, a different unit from the same lot was internally investigated by the manufacturer. This investigation is ongoing. Thermo fisher scientific has initiated the recall process to remove units of ion torrent? Oncomine lung cell-free total nucleic acid (cftna) research assay lot #: 2025404 from the market. Affected customers have been contacted to stop use of the product.
Patient Sequence No: 1, Text Type: N, H10


[129753558] A representative from (b)(6), met with a customer representative from (b)(4) to discuss technical issues regarding validation activities at the (b)(6). (b)(6) complained that a possible contaminant in the ion torrent? Oncomine lung cell-free total nucleic acid (cftna) research assay caused the genetic variant, t790m, to show positive in sample known to be negative for the variant. This was haulting the site's validation activities. No patient samples were being analyzed when the contamination was reported. The product was not used to determine treatment or diagnostic decisions. Ion torrent? Oncomine lung cell-free total nucleic acid (cftna) research assay is labeled "for research use only. Not for use in diagnostic procedures. " the product is not a cleared or approved medical device.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003335080-2018-00002
MDR Report Key8122601
Date Received2018-12-01
Date of Report2018-11-30
Date of Event2018-10-29
Date Mfgr Received2018-10-29
Device Manufacturer Date2018-08-30
Date Added to Maude2018-12-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDR. KELLI TANZELLA
Manufacturer Street7335 EXECUTIVE WAY
Manufacturer CityFREDERICK MD 21704
Manufacturer CountryUS
Manufacturer Postal21704
Manufacturer G1LIFE TECHNOLOGIES CORPORATION
Manufacturer Street7335 EXECUTIVE WAY
Manufacturer CityFREDERICK MD 21704
Manufacturer CountryUS
Manufacturer Postal Code21704
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameION TORRENT? ONCOMINE LUNG CELL-FREE TOTAL NUCLEIC ACID (CFTNA) RESEARCH ASSAY
Generic NameONCOMINE LUNG CELL-FREE TOTAL NUCLEIC ACID (CFTNA) RESEARCH ASSAY
Product CodeOTV
Date Received2018-12-01
Catalog NumberA35864
Lot Number2025404
Device Expiration Date2019-08-30
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerLIFE TECHNOLOGIES CORPORATION
Manufacturer Address7335 EXECUTIVE WAY FREDERICK MD 21704 US 21704


Patients

Patient NumberTreatmentOutcomeDate
10 2018-12-01