MEDLINE RIGID LIGHT HANDLE COVER, STERILE MD-R2

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-12-03 for MEDLINE RIGID LIGHT HANDLE COVER, STERILE MD-R2 manufactured by Aspen Surgical Products, Caledonia.

Event Text Entries

[129313747] Aspen surgical received a report from the distributor that product was found with seal issues. The actual device was returned for evaluation. The manufacturing lot number was available for review. A review of the sample confirmed the reported issue for the light handle being packaged in the seal of the pouch. Analysis of the finished good lot number was reviewed. No non-conformance's were noted during the manufacturing process. According to the manufacturing process, these parts are manually loaded into recessed pockets on a packaging machine per specified work instructions. If incorrectly loaded or not orientated properly, the product interferes with the sealing process. Therefore a likely root cause for the parts in the seal can be attributed to an operator error. Operations and supervisors were notified of this issue during the daily update meetings. Based on this information, no further action is required.
Patient Sequence No: 1, Text Type: N, H10


[129313748] Aspen surgical received a report from the distributor indicating that a rigid light handle cover packaging was found with seal issues. The item was not in use. No injury/death was reported. This report was filed in our complaint handling system as complaint (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1836161-2018-00124
MDR Report Key8124040
Date Received2018-12-03
Date of Report2018-11-05
Date of Event2018-11-05
Date Mfgr Received2018-11-05
Device Manufacturer Date2018-02-20
Date Added to Maude2018-12-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JORDAN HACKERT
Manufacturer Street6945 SOUTHBELT DR. S.E.
Manufacturer CityCALEDONIA MI 49316
Manufacturer CountryUS
Manufacturer Postal49316
Manufacturer Phone6165367508
Manufacturer G1SAME AS ABOVE
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameMEDLINE RIGID LIGHT HANDLE COVER, STERILE
Generic NameLIGHT HANDLE
Product CodeFTA
Date Received2018-12-03
Model NumberMD-R2
Lot Number156457
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerASPEN SURGICAL PRODUCTS, CALEDONIA
Manufacturer Address6945 SOUTHBELT DR. S.E. CALEDONIA MI 49316 US 49316


Patients

Patient NumberTreatmentOutcomeDate
10 2018-12-03

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.