VRV-IIC VACUUM RELIEF VALVE 4004103

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,foreign report with the FDA on 2018-12-03 for VRV-IIC VACUUM RELIEF VALVE 4004103 manufactured by Quest Medical, Inc.

Event Text Entries

[129082326] The alleged incident occurred outside of the us. This medwatch is being submitted because quest distributes a similar device in the us. The complaint sample was evaluated for leakage by connecting it to a water source placed at head height of approximately 4ft in the direction opposite to the flow to replicate static pressure build up. The valve started to leak from under the pressure relief band. The root cause of the reported complaint condition is static pressure build up in the circuit.
Patient Sequence No: 1, Text Type: N, H10


[129082327] The report received states that during a case, blood leakage was seen from below the pressure relief band. The medical staff used the valve to complete the surgery because the amount of blood leaking was small. Additional information received indicates that the incident occurred shortly after the pump was turned off.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1649914-2018-00095
MDR Report Key8125699
Report SourceDISTRIBUTOR,FOREIGN
Date Received2018-12-03
Date of Report2018-12-03
Date Mfgr Received2018-11-09
Date Added to Maude2018-12-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactTOSAN ONOSODE
Manufacturer StreetONE ALLENTOWN PARKWAY
Manufacturer CityALLEN TX 75002
Manufacturer CountryUS
Manufacturer Postal75002
Manufacturer Phone9723326338
Manufacturer G1QUEST MEDICAL, INC
Manufacturer StreetONE ALLENTOWN PARKWAY
Manufacturer CityALLEN TX 75002
Manufacturer CountryUS
Manufacturer Postal Code75002
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVRV-IIC VACUUM RELIEF VALVE
Generic NameCPBP VACUUM RELIEF VALVE
Product CodeDWD
Date Received2018-12-03
Returned To Mfg2018-11-19
Model Number4004103
Catalog Number4004103
Lot Number0537337G03
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerQUEST MEDICAL, INC
Manufacturer AddressONE ALLENTOWN PARKWAY ALLEN TX 75002 US 75002


Patients

Patient NumberTreatmentOutcomeDate
10 2018-12-03

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.