CLINICAL CHEMISTRY CREATININE 03L81-22

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-12-03 for CLINICAL CHEMISTRY CREATININE 03L81-22 manufactured by Abbott Laboratories.

Event Text Entries

[129751950] An evaluation is in process. A follow-up report will be submitted when the evaluation is complete.
Patient Sequence No: 1, Text Type: N, H10


[129751951] The customer observed multiple falsely elevated architect creatinine results from a patient sample. The initial result was 6. 72 mg/dl and repeat results (same instrument) were 6. 72 and 6. 63 mg/dl. The customer uses a reference range of 0. 57-1. 25 mg/dl. As a result, the patient was sent for a redraw and repeat testing was normal (no specific result provided). No other impact to patient management was reported. The customer reported there were no issues with sample collection/handling nor with sample integrity and that other assays were performed on the same sample without any issue. An aliquot of the original sample was sent to labcorp for testing (cobas 710 instrument, creatinine jaffe method) and yielded a result of 2. 646. Additional testing (ex. Bun) correlated with the creatinine redraw normal result, patient history and previously reported results.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1415939-2018-00164
MDR Report Key8126516
Date Received2018-12-03
Date of Report2019-01-14
Date of Event2018-11-06
Date Mfgr Received2019-01-14
Date Added to Maude2018-12-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224667-512
Manufacturer G1ABBOTT LABORATORIES
Manufacturer Street100 ABBOTT PARK ROAD
Manufacturer CityABBOTT PARK IL 600643500
Manufacturer CountryUS
Manufacturer Postal Code600643500
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCLINICAL CHEMISTRY CREATININE
Generic NameCREATININE
Product CodeCGX
Date Received2018-12-03
Catalog Number03L81-22
Lot Number75410UN17
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerABBOTT LABORATORIES
Manufacturer Address100 ABBOTT PARK ROAD ABBOTT PARK IL 600643500 US 600643500


Patients

Patient NumberTreatmentOutcomeDate
10 2018-12-03

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.