ELECSYS FERRITIN 03737551190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-12-04 for ELECSYS FERRITIN 03737551190 manufactured by Roche Diagnostics.

Event Text Entries

[129518430] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[129518431] The customer questioned results for 1 neonatal patient tested for elecsys ferritin (ferritin) on a cobas 8000 e 602 module. Thirty minutes after being born on (b)(6) 2018, the patient presented with neurological degradation with hypotonia, severe hypoglycemia and hepatocellular insufficiency. At 12:25 a. M. On (b)(6) 2018 the patient was tested for ferritin and the result was 57,907 ug/l suggesting hyperferritinemia. The evening of (b)(6) 2018 biotin treatment was started based on a potential metabolic disease. On (b)(6) 2018 at 5:30 a. M. A new sample was obtained and the ferritin result had decreased to 90 ng/l. The rest of the test results were consistent with the results from (b)(6) 2018. The customer suspects an interference with biotin affecting the ferritin results. There was no allegation that an adverse event occurred. The e602 module serial number was (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-04621
MDR Report Key8129707
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-12-04
Date of Report2018-12-19
Date of Event2018-09-25
Date Mfgr Received2018-11-13
Date Added to Maude2018-12-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameELECSYS FERRITIN
Generic NameRADIOIMMUNOASSAY (TWO-SITE SOLID PHASE), FERRITIN
Product CodeJMG
Date Received2018-12-04
Model NumberNA
Catalog Number03737551190
Lot NumberASKU
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-12-04

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.