COMPELLA CAP US SCALE PD PPM P7800A000001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-12-04 for COMPELLA CAP US SCALE PD PPM P7800A000001 manufactured by Hill-rom Batesville.

Event Text Entries

[129589741] The hill-rom technician found the patient control board assembly needed to be replaced. It is necessary for the compella? Bariatric bed system to have an effective maintenance program. We recommend that you do annual preventive maintenance.?? Specified checks:?? All caregiver and patient controls, along with their indicators, operate correctly; this includes controls for bed articulation, width extend and retract, scale, bed exit, surface functions, nurse call, cpr, foot longer and shorter, pendant controls, etc. Note: the communication cable that connects the asu to the bed frame must be fully connected and the four thumbscrews tightened for the powered width controls to operate. The motors operate correctly: during articulation, there is no friction, abnormal noises, or an audible indication of overload. A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in 2018. It is unknown if the facility performed any other preventative maintenance on this bed. The technician replaced the patient control board assembly to resolve the issue. Based on this information, no further action is required.
Patient Sequence No: 1, Text Type: N, H10


[129589743] Hill-rom received a report from the account stating the head of the bed was going up on its own. The bed was located at the account. There was no patient/user injury reported. This report was filed in our complaint handling system as complaint #(b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1824206-2018-00464
MDR Report Key8130267
Date Received2018-12-04
Date of Report2018-11-15
Date of Event2018-11-15
Date Mfgr Received2018-11-15
Device Manufacturer Date2016-06-27
Date Added to Maude2018-12-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactEMILY MITCHELL
Manufacturer Street1069 STATE ROUTE 46 EAST
Manufacturer CityBATESVILLE IN 47006
Manufacturer CountryUS
Manufacturer Postal47006
Manufacturer Phone8129310130
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCOMPELLA CAP US SCALE PD PPM
Generic NameBARIATRIC BED
Product CodeOSI
Date Received2018-12-04
Model NumberP7800A000001
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHILL-ROM BATESVILLE
Manufacturer Address1069 STATE ROUTE 46 EAST BATESVILLE IN 47006 US 47006


Patients

Patient NumberTreatmentOutcomeDate
10 2018-12-04