HANA TABLE 6875

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2018-12-05 for HANA TABLE 6875 manufactured by Mizuho Orthopedic Systems, Inc..

Event Text Entries

[129316707] Preventive maintenance was completed on 08/02/2018 and 11/14/2018 and the table passed the inspection. Conclusion of investigation is operator error of not having the patient safely strapped to the table and removing the perineal post prematurely before transferring patient. Two in-service were provided 11/15/2018 & 11/20/2018 to ensure that all staff members were in-serviced. The owner's manual includes instructions for securing the patient with a safety strap.
Patient Sequence No: 1, Text Type: N, H10


[129316708] Patient fell off table - outcome unknown. As reported, the patient safety strap was not used. When preparing for patient transfer, the perineal post was pulled prematurely and the patient fell.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2921578-2018-00035
MDR Report Key8132818
Report SourceCOMPANY REPRESENTATIVE
Date Received2018-12-05
Date of Report2018-12-05
Date of Event2018-11-08
Date Mfgr Received2018-11-09
Device Manufacturer Date2016-11-04
Date Added to Maude2018-12-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactANNE LEBLANC
Manufacturer Street30031 AHERN AVENUE
Manufacturer CityUNION CITY CA 945871234
Manufacturer CountryUS
Manufacturer Postal945871234
Manufacturer Phone5104291500
Manufacturer G1MIZUHO ORTHOPEDIC SYSTEMS, INC.
Manufacturer Street30031 AHERN AVENUE
Manufacturer CityUNION CITY CA 945871234
Manufacturer CountryUS
Manufacturer Postal Code945871234
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHANA TABLE
Generic NameSURGICAL TABLE
Product CodeJEA
Date Received2018-12-05
Model Number6875
Catalog Number6875
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMIZUHO ORTHOPEDIC SYSTEMS, INC.
Manufacturer Address30031 AHERN AVENUE UNION CITY CA 945871234 US 945871234


Patients

Patient NumberTreatmentOutcomeDate
10 2018-12-05