VIDAS? BRAHMS PROCALCITONIN 30450

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-12-05 for VIDAS? BRAHMS PROCALCITONIN 30450 manufactured by Biomerieux Sa.

Event Text Entries

[130061252] A customer in (b)(6) reported a falsely underestimated result in association with the vidas? Brahms procalcitonin assay. The customer stated the initial test did not give the correct result based on the background reading, rfv, and patient condition, and that the repeat test was consistent with the patient's critical condition with sepsis. The patient had a positive blood culture on (b)(6) 2018, and pct was run on (b)(6) 2018. The customer mentioned they found a stuck solid phase receptacles (spr? ) at the end of the initial run. Position a1 - initial test: background reading : 23, sample rfv: -9, result (ng/ml): < 0. 05. Position b1 - repeat test: background reading : 169, sample rfv: 763, result (ng/ml): 15. 12. The customer reported that there was no impact to patient treatment as the repeat result which was consistent with the blood culture positive result, was reported to the physician. It was reported that the patient condition improved and they were discharged. A biom? Rieux internal investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2018-00250
MDR Report Key8133944
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-12-05
Date of Report2019-02-13
Date of Event2018-09-04
Date Mfgr Received2019-01-15
Device Manufacturer Date2018-04-05
Date Added to Maude2018-12-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JEFF SCANLAN
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147318694
Manufacturer G1BIOMERIEUX SA
Manufacturer StreetCHEMIN DE L ORME
Manufacturer CityMARCY L ETOILE, RHONE 69280,
Manufacturer CountryFR
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVIDAS? BRAHMS PROCALCITONIN
Generic NameVIDAS? BRAHMS PCT
Product CodePRI
Date Received2018-12-05
Catalog Number30450
Lot Number1006397040
Device Expiration Date2019-03-21
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX SA
Manufacturer AddressCHEMIN DE L ORME MARCY L ETOILE, RHONE 69280, FR


Patients

Patient NumberTreatmentOutcomeDate
10 2018-12-05

© 2026 FDA.report