ARCHITECT SYPHILIS TP 08D06-39

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2018-12-06 for ARCHITECT SYPHILIS TP 08D06-39 manufactured by Abbott Germany.

Event Text Entries

[130620221] This report is being filed on an international product. Ln 8d06-39 that has a similar product in the us ln 8d06-31 and 8d06-41. Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, review of instrument logs, a review of labeling and sensitivity testing on lots 88489li00 and 89366li00. No adverse trend was identified for the customer's issue. Labeling was reviewed and found to be adequate. Both clinical seroconversion panels and in-house sensitivity testing met all specifications. No file kit testing could be performed for lot 82020li00 as this lot expired as of 2018-09-25. Return patient samples in (b)(6) are not allowed to be exported out of country due to (b)(6) government restrictions. Based on all available information and abbott diagnostics' complaint investigation no product deficiency was identified. This event has been reported on 3 separate suspect medical devices, please also see the following manufacturer report numbers 3002809144-2018-00343 for lot 82020li00; 3002809144-2018-00541 for lot 89366li00.
Patient Sequence No: 1, Text Type: N, H10


[130620222] The customer observed a (b)(6) result on the architect i2000sr analyzer. The following data was provided sid (b)(6): initial (b)(6). Tppa (b)(6) and j&j (b)(6). There was no impact to patient management reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002809144-2018-00540
MDR Report Key8139958
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2018-12-06
Date of Report2018-12-06
Date Mfgr Received2018-11-13
Device Manufacturer Date2018-05-29
Date Added to Maude2018-12-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224667-512
Manufacturer G1ABBOTT GERMANY
Manufacturer StreetMAX-PLANCK-RING 2
Manufacturer CityWIESBADEN 65205
Manufacturer CountryGM
Manufacturer Postal Code65205
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameARCHITECT SYPHILIS TP
Generic NameSYPHILIS
Product CodeMTN
Date Received2018-12-06
Catalog Number08D06-39
Lot Number88489LI00
Device Expiration Date2019-04-23
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT GERMANY
Manufacturer AddressMAX-PLANCK-RING 2 WIESBADEN 65205 GM 65205


Patients

Patient NumberTreatmentOutcomeDate
10 2018-12-06

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.