YASARGIL TI PERM STD-CLIPSTRG-CVD7.4MM FT823T

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2018-12-07 for YASARGIL TI PERM STD-CLIPSTRG-CVD7.4MM FT823T manufactured by Aesculap Ag.

Event Text Entries

[130037469] (b)(4). Manufacturing site evaluation: no product was returned. Batch nor serial number provided. A review of the device quality and manufacturing history records was not possible. Conclusion- failure is possible related to insufficient usage. Rationale - it is very important to ensure that the occlusion of the aneurysm or vessel is achieved and that the clip sits correctly on the aneurysm neck during and after the implantation and sits tight on the blood vessel. According to the ifu, the instructions must be used to avoid damage or a failure of the implant.
Patient Sequence No: 1, Text Type: N, H10


[130037470] It was reported that there was postoperative slippage of a yasargil aneurysm clip. The initial surgery and implantation of the clip occurred on (b)(6) 2009. It was noted that another manufacturer's clip had also been used near the yasargil clip. On (b)(6) 2009, the slip-out occurred. Limited information was provided, but additional data has been requested.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610612-2018-00576
MDR Report Key8142146
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2018-12-07
Date of Report2018-12-07
Date of Event2009-06-19
Date Facility Aware2018-11-28
Date Mfgr Received2018-11-14
Date Added to Maude2018-12-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. NICOLE BROYLES
Manufacturer Street615 LAMBERT POINTE DRIVE
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3145515988
Manufacturer G1AESCULAP AG
Manufacturer StreetPO BOX 40
Manufacturer CityTUTTLINGEN, 78501
Manufacturer CountryGM
Manufacturer Postal Code78501
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameYASARGIL TI PERM STD-CLIPSTRG-CVD7.4MM
Generic NameCEREBRO VASCULAR CLIPS
Product CodeHCH
Date Received2018-12-07
Model NumberFT823T
Catalog NumberFT823T
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerAESCULAP AG
Manufacturer AddressPO BOX 40 TUTTLINGEN, 78501 GM 78501


Patients

Patient NumberTreatmentOutcomeDate
10 2018-12-07

© 2024 FDA.report