ELECSYS TOTAL PSA IMMUNOASSAY 08791686160

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2018-12-03 for ELECSYS TOTAL PSA IMMUNOASSAY 08791686160 manufactured by Roche Diagnostics.

Event Text Entries

[130050483]
Patient Sequence No: 1, Text Type: N, H10


[130050484] The customer complained of questionable elecsys free psa immunoassay and elecsys total psa immunoassay results for 1 patient sample tested on a cobas e 411 immunoassay analyzer serial number (b)(4). The customer questioned a possible interference in the patient sample. This medwatch will cover the total psa data. Please refer to the medwatch with patient identifier (b)(6) for data on the free psa. The initial total psa result was 0. 063 ng/ml with a repeat result of 0. 063 ng/ml. The initial free psa result was 0. 151 ng/ml with repeat results of 0. 154 ng/ml and 0. 156 ng/ml. No erroneous results were reported outside of the laboratory. There was no adverse event. The field service engineer (fse) performed an instrument performance check that was acceptable. The instrument was performing within specifications. The investigation is currently ongoing.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-04726
MDR Report Key8147426
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2018-12-03
Date of Report2019-01-03
Date of Event2018-11-09
Date Mfgr Received2018-11-20
Date Added to Maude2018-12-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameELECSYS TOTAL PSA IMMUNOASSAY
Generic NameTOTAL,PROSTATE SPECIFIC ANTIGEN(NONCOMPLEXED&COMPLEXED) FOR DETECTION OF PROS
Product CodeMTF
Date Received2018-12-03
Model NumberNA
Catalog Number08791686160
Lot Number00315719
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-12-03

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.