FUSE FULL SPECTRUM ENDOSCOPY SYSTEM FSC330-A6334 60093

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2018-12-10 for FUSE FULL SPECTRUM ENDOSCOPY SYSTEM FSC330-A6334 60093 manufactured by Boston Scientific Corporation.

Event Text Entries

[130049422] (b)(4). A fuse 1c colonoscope was returned for analysis. A functional evaluation was performed and found a distorted center image. Additionally, corrosion and signs of past fluid invasion were found on the electrical connector. The reported issue was confirmed. The device was not properly cleaned after being used. Fluid was present in the electrical connector that caused the corrosion. This interfered with the imaging signals resulting in the intermittent imaging issue. Therefore, the most probable root cause is cause traced to maintenance.
Patient Sequence No: 1, Text Type: N, H10


[130049423] It was reported to boston scientific corporation that a fuse 1c colonoscope was used during a colonoscopy performed on (b)(6) 2018. According to the complainant, during the procedure, the center image went in and out. The anatomy of the patient was not visible when the center image went in and out. Another fuse 1c colonoscope was used to complete the procedure. There were no patient complications reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2018-61869
MDR Report Key8147978
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2018-12-10
Date of Report2018-12-10
Date of Event2018-11-12
Date Mfgr Received2018-11-12
Device Manufacturer Date2015-11-03
Date Added to Maude2018-12-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCAROLE MORLEY
Manufacturer Street300 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834015
Manufacturer G1ENDOCHOICE, INC.
Manufacturer Street11810 WILLS ROAD
Manufacturer CityALPHARETTA GA 30009
Manufacturer CountryUS
Manufacturer Postal Code30009
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameFUSE FULL SPECTRUM ENDOSCOPY SYSTEM
Generic NameENDOSCOPE, ACCESSORIES, IMAGE POST-PROCESSING FOR COLOR ENHANCEMENT
Product CodePEA
Date Received2018-12-10
Returned To Mfg2018-11-12
Model NumberFSC330-A6334
Catalog Number60093
Lot Number0000700424
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer Address300 BOSTON SCIENTIFIC WAY MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
10 2018-12-10

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.