[130316119] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[130316120] A customer reported using expired cidex? Opa solution past the 75-day open date of the bottle and the instruments that were cleaned in the solution were released and used on patients. They stated the solution was tested with the cidex? Opa test strips for the minimum effective concentration (mec) and passed with each use and the solution was heated per instructions for use (ifu) to 20? C (68? F). There was no report of infection, injury or harm to patient(s) associated with this issue. The instructions for use (ifu) states that once the bottle has been opened, the solution remaining in the container may be stored for up to 75 days until used. The customer was advised to always follow the ifu and to discard any solution passed the expiration date. As a matter of policy, advanced sterilization products (asp) has decided to report cases where a customer uses expired product and releases it for use on patients since they cannot guarantee high-level disinfection has been achieved.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2084725-2018-00780
MDR Report Key8151656
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2018-12-11
Date of Report2018-11-21
Date of Event2018-11-21
Date Mfgr Received2019-01-10
Device Manufacturer Date2018-02-14
Date Added to Maude2018-12-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. GABRIEL ALFAGEME
Manufacturer Street33 TECHNOLOGY DRIVE
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal92618
Manufacturer G1ADVANCED STERILIZATION PRODUCTS
Manufacturer Street33 TECHNOLOGY DRIVE
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal Code92618
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCIDEX? OPA SOLUTION
Generic NameBIOCIDES SOLUTIONS
Product CodeMED
Date Received2018-12-11
Catalog Number20390
Lot Number805000013
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerADVANCED STERILIZATION PRODUCTS
Manufacturer Address33 TECHNOLOGY DRIVE IRVINE CA 92618 US 92618


Patients

Patient NumberTreatmentOutcomeDate
10 2018-12-11

© 2024 FDA.report