[130498287] Dhr - no deviations found in the production process or in the finished goods inspection. The products were released without quality issues. Retained samples were pulled out one for analysis. Full specification testing was done and all parameters were found to be acceptable. Appearance is good and no sticky issue was observed. From the investigation, the complaint was not confirmed. Root cause can not be determined as there was no sample received for evaluation and the complaint was not confirmed. However, this problem can occurred due to several factors that affect the performance of the product during the preparation and application, like water temperature, rolling technique, excess water removal, rubbing technique and drying time. With regards to sticky issue, the product must be stored within the indicated 15oc- 30oc. Storage outside the indicated temperature might cause the migration of the resin that result to stickiness.
Patient Sequence No: 1, Text Type: N, H10


[130498288] 2 of 2 reports - other mfg reports number: 9680091-2018-00039. It was reported that the facility is having issues with "sticky" tcc-ez kits that won't harden properly. Additional information has been requested.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9680091-2018-00040
MDR Report Key8157532
Report SourceUSER FACILITY
Date Received2018-12-12
Date of Report2018-11-19
Date Mfgr Received2018-11-19
Date Added to Maude2018-12-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactUSER VIVIAN NELSON
Manufacturer Street311 ENTERPRISE DRIVE
Manufacturer CityPLAINSBORO NJ 08536
Manufacturer CountryUS
Manufacturer Postal08536
Manufacturer Phone6099362393
Manufacturer G1DERMA SCIENCES, INC.
Manufacturer Street104 SHORTING ROAD
Manufacturer CityTORONTO, ONTARIO M1S 3S4
Manufacturer CountryCA
Manufacturer Postal CodeM1S 3S4
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTCC-EZ CASTING SYSTEM 3" SINGLE APPLICATION
Generic NameTCC-EZ
Product CodeITG
Date Received2018-12-12
Catalog NumberTCC23002
Lot Number0918021
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerDERMA SCIENCES, INC.
Manufacturer Address104 SHORTING ROAD 104 SHORTING ROAD TORONTO, ONTARIO M1S 3S4 CA M1S 3S4


Patients

Patient NumberTreatmentOutcomeDate
10 2018-12-12