RX LOCKING DEVICE AND BIOPSY CAP M00545260 4526S

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2018-12-13 for RX LOCKING DEVICE AND BIOPSY CAP M00545260 4526S manufactured by Boston Scientific Corporation.

Event Text Entries

[130230570] (b)(4). Although expected, the device has not been received for analysis. Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
Patient Sequence No: 1, Text Type: N, H10


[130230571] It was reported to boston scientific corporation that an rx locking device biopsy cap was used during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2018. According to the complainant, the sponge in the biopsy cap detached and became stuck in the channel of the scope during procedure. It was reported that water soluble lubricant was applied to the top of the biopsy cap prior to use. The detached piece was successfully removed from the scope using hemostats. The procedure was completed using another rx locking device biopsy cap. There was no serious injury nor adverse patient effects reported as a result of this event. The patient's condition at the conclusion of the procedure was reported to be stable.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2018-61969
MDR Report Key8159983
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2018-12-13
Date of Report2019-02-05
Date of Event2018-11-12
Date Mfgr Received2019-01-11
Date Added to Maude2018-12-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCAROLE MORLEY
Manufacturer Street300 BOSTON SCIENTIFIC WAY ,
Manufacturer CityMARLBOROUGH MA 017521242
Manufacturer CountryUS
Manufacturer Postal017521242
Manufacturer Phone5086834015
Manufacturer G1BOSTON SCIENTIFIC CORPORATION
Manufacturer Street780 BROOKSIDE DRIVE ,
Manufacturer CitySPENCER IN 47460
Manufacturer CountryUS
Manufacturer Postal Code47460
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRX LOCKING DEVICE AND BIOPSY CAP
Generic NameENDOSCOPE AND/OR ACCESSORIES
Product CodeODC
Date Received2018-12-13
Returned To Mfg2018-12-17
Model NumberM00545260
Catalog Number4526S
Lot Number0022601798
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer Address300 BOSTON SCIENTIFIC WAY , MARLBOROUGH MA 017521242 US 017521242


Patients

Patient NumberTreatmentOutcomeDate
10 2018-12-13

© 2024 FDA.report