KEY SURGICAL UNIVERSAL LIGHT HANDLE COVER LHC-201-160

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2018-12-14 for KEY SURGICAL UNIVERSAL LIGHT HANDLE COVER LHC-201-160 manufactured by Key Surgical Llc.

Event Text Entries

[130487607] The event was reported to key surgical by (b)(4). The reported event occurred in (b)(6); however, key surgical has historically marketed similar versions of the device in the united states using the same part number. The device was unavailable for investigation. Key surgical, together with (b)(4) requested additional information from the hospital facility that initially reported the complaint information to the distributor, (b)(4) on several separate dates. The user facility was unable to provide further information on events leading up to the described event, but was able to confirm part number, lot number and device description information. The cause of the reported malfunction/adverse event is unknown. The lot history information including device history records, inspection records and requests for information on supplier manufacturing information for the reported lot resulted in no non-conformities, abnormalities or information related to the complaint root cause. Through 12/12/2018, review of complaint history against sales data reveals that the complaint rate for the light handle cover t is (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[130487608] Surgical light handle cover placed over light handle prior to surgical start. Intraoperatively, the surgeon reached for the light to change the angle. The light handle cover split and fell into the surgical field. The user facility reported that the patient was treated with antibiotics postoperatively. No other patient impact was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2183785-2018-00002
MDR Report Key8166241
Report SourceDISTRIBUTOR
Date Received2018-12-14
Date of Report2018-12-13
Date of Event2018-10-22
Date Mfgr Received2018-11-15
Device Manufacturer Date2018-04-04
Date Added to Maude2018-12-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAKE SCHULTZ
Manufacturer Street8101 WALLACE ROAD
Manufacturer CityEDEN PRAIRIE MN 55344
Manufacturer CountryUS
Manufacturer Postal55344
Manufacturer Phone8005417995
Manufacturer G1KEY SURGICAL LLC
Manufacturer Street8101 WALLACE ROAD
Manufacturer CityEDEN PRAIRIE MN 55344
Manufacturer CountryUS
Manufacturer Postal Code55344
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameKEY SURGICAL UNIVERSAL LIGHT HANDLE COVER
Generic NameLIGHT HANDLE COVER
Product CodeFTA
Date Received2018-12-14
Model NumberLHC-201-160
Lot Number64055
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerKEY SURGICAL LLC
Manufacturer Address8101 WALLACE ROAD EDEN PRAIRIE MN 55344 US 55344


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2018-12-14

© 2026 FDA.report