[130635933] The field for sections was not populated due to a technical limitation: common name: catheter, peritoneal; product code: gbw. This mdr is being filed after the associated complaint was reviewed under remediation protocol (b)(4), complaint/mdr retrospective review and remediation and reassessed as reportable. Additional complaint investigation and record remediation was not performed.
Patient Sequence No: 1, Text Type: N, H10


[130635934] It was reported that during the peritoneal dialysis, a luer lock connection on tenckhoff acute peritoneal dialysis catheter was leaking when connected o another manufacturer's dialysis bag. The luer lock was not a "snug fit" with the other manufacturer's kit. There were no adverse effects reported due to the event and no report of an additional procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1820334-2018-03821
MDR Report Key8167876
Date Received2018-12-16
Date of Report2018-12-16
Date Mfgr Received2018-12-13
Device Manufacturer Date2014-07-21
Date Added to Maude2018-12-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. LARRY POOL
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal47404
Manufacturer Phone8128294891
Manufacturer G1COOK INC
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal Code47404
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameTENCKHOFF ACUTE PERITONEAL DIALYSIS CATHETER SET
Product CodeGBW
Date Received2018-12-16
Model NumberNA
Catalog NumberC-TAPS-9.5-31-12-0-10.5
Lot Number5112104
Device Expiration Date2017-07-21
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCOOK INC
Manufacturer Address750 DANIELS WAY BLOOMINGTON IN 47404 US 47404


Patients

Patient NumberTreatmentOutcomeDate
10 2018-12-16