CANON CAT-870B

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-12-17 for CANON CAT-870B manufactured by Canon Medical Systems Corporation.

Event Text Entries

[131381504] Table tilted uncommanded into c- arm of system causing damage to the internal workings of the table. The patient was on the table when this occurred, but was unhurt. It was reported by the customer that the table began tilting with the patient on the table without operation intervention. They were able to get the patient off the table without injury. However, the table continued to tilt into the c-arm causing irreparable damage to the table. A replacement table has been ordered. Please note that prior to the table tilting, a nurse pierced a heparin bag that was to be hung on the iv pole on the left side of the table. The bag ruptured saturating the tilt console and the foot of the table top with the heparin. The areas were cleaned, but soon after the event (table tilt) occurred. The table, table control powerboard and connecting cable are being returned to the manufacturer for investigation.
Patient Sequence No: 1, Text Type: N, H10


[131381505] Table tilted uncommanded into c-arm of system causing damage to the internal workings of the table. Patient was on the table, but was unhurt.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2020563-2018-00003
MDR Report Key8170451
Date Received2018-12-17
Date of Report2018-12-10
Date of Event2018-11-08
Date Mfgr Received2018-11-09
Device Manufacturer Date2008-05-01
Date Added to Maude2018-12-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. PAUL BIGGINS
Manufacturer Street2441 MICHELLE DRIVE
Manufacturer CityTUSTIN CA 92780
Manufacturer CountryUS
Manufacturer Postal92780
Manufacturer Phone7147305000
Manufacturer G1CANON MEDICAL SYSTEMS CORPORATION
Manufacturer Street1385 SHIMOISHIGAMI
Manufacturer CityOTAWARA-SHI, TOCHIGI 324-8550
Manufacturer CountryJA
Manufacturer Postal Code324-8550
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCANON
Generic NameCATHETERIZATION TABLE
Product CodeIXR
Date Received2018-12-17
Model NumberCAT-870B
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCANON MEDICAL SYSTEMS CORPORATION
Manufacturer Address1385 SHIMOISHIGAMI OTAWARA-SHI, TOCHIGI 324-8550 JA 324-8550


Patients

Patient NumberTreatmentOutcomeDate
10 2018-12-17

© 2024 FDA.report