AMPLATZER DUCT OCCLUDER 9-PDA-003

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2018-12-18 for AMPLATZER DUCT OCCLUDER 9-PDA-003 manufactured by Aga Medical Corporation.

Event Text Entries

[130626983] An event of embolization was reported. The results of the investigation are inconclusive since the device was not returned for analysis. Based on the information received, the cause of the reported incident could not be conclusively determined. Information from the field indicated that " a larger amplatzer duct occluder was used to close the pda. "
Patient Sequence No: 1, Text Type: N, H10


[130626984] On (b)(6) 2018, a 5/4mm amplatzer duct occluder was implanted. Three hours after the procedure was completed and the patient was recovering in the nicu, the device embolized to the descending aorta. A 5f bioptome was used to grab the pa side pin (end screw) and the device was smoothly resheathed and removed. A larger amplatzer duct occluder was used to close the pda. There were no patient consequences reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2135147-2018-00241
MDR Report Key8173386
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2018-12-18
Date of Report2018-12-18
Date of Event2018-11-02
Date Mfgr Received2018-11-20
Device Manufacturer Date2018-09-12
Date Added to Maude2018-12-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTEPHANIE O' SULLIVAN
Manufacturer Street5050 NATHAN LANE NORTH
Manufacturer CityPLYMOUTH MN 55442
Manufacturer CountryUS
Manufacturer Postal55442
Manufacturer Phone6517565400
Manufacturer G1AGA MEDICAL CORPORATION
Manufacturer Street5050 NATHAN LANE NORTH
Manufacturer CityPLYMOUTH MN 55442
Manufacturer CountryUS
Manufacturer Postal Code55442
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAMPLATZER DUCT OCCLUDER
Generic NameOCCLUDER, PATENT DUCTUS, ARTERIOSUS
Product CodeMAE
Date Received2018-12-18
Model Number9-PDA-003
Catalog Number9-PDA-003
Lot Number6631855
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerAGA MEDICAL CORPORATION
Manufacturer Address5050 NATHAN LANE NORTH PLYMOUTH MN 55442 US 55442


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Required No Informationntervention 2018-12-18