AUTOCLAVABLE INTERNAL HANDLES 8011-0501-01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-12-18 for AUTOCLAVABLE INTERNAL HANDLES 8011-0501-01 manufactured by Zoll Medical Corporation.

Event Text Entries

[130648713] After cardio pulmonary bypass defibrillation was needed using internal paddles set at 5 joules. There was a delay between pushing the button and delivering joules. The first time nothing was delivered. The button was pushed three more times at 5 joules and only 1 joule was actually delivered. Was in the process of switching to another set of defibrillator handles when the patient converted. Reported issue was investigated, it was determined that the internal paddles were 17 years old and passed its sterilization cycle threshold of 100 sterilizations. Manufacturer response for defibrillator paddles, zoll (per site reporter). Paddles were beyond recommended age. Zoll provided ordering information for new paddles.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8173754
MDR Report Key8173754
Date Received2018-12-18
Date of Report2018-12-13
Date of Event2018-07-23
Report Date2018-12-13
Date Reported to FDA2018-12-13
Date Reported to Mfgr2018-12-18
Date Added to Maude2018-12-18
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAUTOCLAVABLE INTERNAL HANDLES
Generic NameDC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES)
Product CodeLDD
Date Received2018-12-18
Catalog Number8011-0501-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Age17 YR
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerZOLL MEDICAL CORPORATION
Manufacturer Address269 MILL ROAD CHELMSFORD MA 01824 US 01824


Patients

Patient NumberTreatmentOutcomeDate
10 2018-12-18