COOK FLEXIBLE MYOCARDIAL BIOPSY FORCEPS FBF-5.2-120

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-12-18 for COOK FLEXIBLE MYOCARDIAL BIOPSY FORCEPS FBF-5.2-120 manufactured by Cook Inc.

Event Text Entries

[131143120] Common name: device, biopsy, endomyocardial. Product code: dwz. This mdr is being filed after the associated complaint was reviewed under remediation protocol (b)(4), complaint/mdr retrospective review and remediation and reassessed as reportable. Additional complaint investigation and record remediation was not performed.
Patient Sequence No: 1, Text Type: N, H10


[131143121] It was reported that during the biopsy of left ventricle, the flexible myocardial biopsy forceps would not open inside the patient when advanced through the guiding catheter. After the procedure, the user checked if the device shaft of the device was bent. It was further noted that the handle would not actuate the forceps due to a missing rivet and internal corrosion. The procedure was completed with another device, and there were no injuries or additional procedures reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1820334-2018-03842
MDR Report Key8173808
Date Received2018-12-18
Date of Report2018-12-18
Date of Event2017-01-12
Date Mfgr Received2018-12-13
Device Manufacturer Date2016-01-13
Date Added to Maude2018-12-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. LARRY POOL
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal47404
Manufacturer Phone8128294891
Manufacturer G1COOK INC
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal Code47404
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCOOK FLEXIBLE MYOCARDIAL BIOPSY FORCEPS
Product CodeDWZ
Date Received2018-12-18
Returned To Mfg2017-01-26
Model NumberNA
Catalog NumberFBF-5.2-120
Lot Number6515681
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCOOK INC
Manufacturer Address750 DANIELS WAY BLOOMINGTON IN 47404 US 47404


Patients

Patient NumberTreatmentOutcomeDate
10 2018-12-18

© 2026 FDA.report