UL PRO FUSED HEADLIGHT CABLE 9FT, 275CM 001388LX9

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-12-18 for UL PRO FUSED HEADLIGHT CABLE 9FT, 275CM 001388LX9 manufactured by Integra York, Pa Inc..

Event Text Entries

[130986210] The device was not returned to the manufacturer for analysis. The plant investigation is in progress and a supplemental medwatch report will be submitted upon completion of the investigation.
Patient Sequence No: 1, Text Type: N, H10


[130986211] A biomed tech reported to integra that on (b)(6) 2018, during the setup for a case, the ul pro fused headlight cable 9ft, 275 cm become very dim while the surgeon was using the device for roughly 10 minutes. Upon further checking, the lightsource malfunctioned and become too hot, which burned and cracked the end of the lens of the cable. The patient was prepped for surgery, but the device was not in contact with the patient. There was no patient injury/death alleged, and no medical intervention was required. Initially it was reported of an unspecified delay in surgery due to product problem. Additional information was received on (b)(6) 2018, and reported that there was no delay in the procedure due to the product problem, and no patient impact.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2523190-2018-00179
MDR Report Key8175605
Date Received2018-12-18
Date of Report2018-12-04
Date of Event2018-11-30
Date Mfgr Received2018-12-28
Date Added to Maude2018-12-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactUSER KIMBERLY SHELLY
Manufacturer Street311 ENTERPRISE DRIVE
Manufacturer CityPLAINSBORO NJ 08536
Manufacturer CountryUS
Manufacturer Postal08536
Manufacturer Phone6099362393
Manufacturer G1INTEGRA YORK, PA INC.
Manufacturer Street589 DAVIES DRIVE
Manufacturer CityYORK PA 17402
Manufacturer CountryUS
Manufacturer Postal Code17402
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameUL PRO FUSED HEADLIGHT CABLE 9FT, 275CM
Generic NameCABLES
Product CodeFST
Date Received2018-12-18
Catalog Number001388LX9
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerINTEGRA YORK, PA INC.
Manufacturer Address589 DAVIES DRIVE 589 DAVIES DRIVE YORK PA 17402 US 17402


Patients

Patient NumberTreatmentOutcomeDate
10 2018-12-18

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.