FREEDOM EVO 150 10641150

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-12-19 for FREEDOM EVO 150 10641150 manufactured by Tecan Schweiz Ag.

Event Text Entries

[131800188] A tecan field service representative (fse) was dispatched to evaluate the report of the front panel falling. The fse replaced the gas spring as it was past the expected lifetime (i. E. 3 years). During the last routine preventive maintenance intervention, the gas spring was not replaced in error. A capa has been opened to investigate the issue. The evo 150 front panels are made of plexiglass and are light (<4. 5kg) and have been assessed in the past to not cause a serious injury. However, in an abundance of caution, due to prior events on larger and heavier, though similar models that resulted in moderate injuries (and for which mdrs were filed), a mdr is being filed.
Patient Sequence No: 1, Text Type: N, H10


[131800189] The laboratory called to report that one of the door hinges is starting to fail and the front panel hit the operator on the head. There were no injuries (e. G. Bumps, bruises, cuts, etc) and no medical attention required. The problem started about a week ago and seems to get worse as we open and close the door (front panel).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003402518-2018-00003
MDR Report Key8178760
Date Received2018-12-19
Date of Report2018-12-19
Date of Event2018-07-17
Date Mfgr Received2018-07-17
Device Manufacturer Date2013-11-13
Date Added to Maude2018-12-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. JILLIAN WALKER
Manufacturer StreetSEESTRASSE 103
Manufacturer CityMAENNEDORF, ZH 8708
Manufacturer CountrySZ
Manufacturer Postal8708
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameFREEDOM EVO 150
Generic NamePIPETTING STATION FOR CLINICAL USE
Product CodeJQW
Date Received2018-12-19
Model Number10641150
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTECAN SCHWEIZ AG
Manufacturer AddressSEESTRASSE 103 MAENNEDORF, ZH 8708 SZ 8708


Patients

Patient NumberTreatmentOutcomeDate
10 2018-12-19

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.